> [Edit: Ah - got it! It’s measuring the frequency of the very lowest points - the ‘every second dip’]
So that's the danger of [the authors] just reading the number off the screen without understanding what the scope is showing.To measure frequency, typically the scope will measure the max and min of a waveform, then select a point halfway between them as the trigger point, then measure the time between successive triggers. Also, the trigger usually has an intentional hysteresis, so that noise near the midpoint does not cause multiple triggers on a slowly-rising waveform.
The little dips don't go down far enough below the midpoint to activate the hysteresis window, which is why it only counts every second dip.
Edit to add: Also, while it's nice that they show all the waveforms with the same vertical scale, they really need to adjust the vertical attenuation for the 0% and 25% brightness steps if they want to measure frequency by reading a number off the screen. Auto measurements won't work in these cases, when the waveform fills less than half a vertical division.
And the waveform trigger is also set incorrectly (0V, with DC coupling on the input...) because none of the waveforms have a rising edge on the center graticule line, where the trigger annunciator is.